Highlights from this day
  • bac_water_bill — net peptide is gross mass minus water minus counterion. that is why 11.4mg of powder can be a correct 10mg vial 17:00
  • lisbon_lot — if you only read one line on a certificate, read the lot number. if you read two, read the assay, take that with a pinch of salt 18:18
  • chain_of_custody — tfa counterion runs higher by mass than acetate. if the certificate does not say, you cannot do the mass balance 18:45
  • chain_of_custody — water content is why an honest 10mg vial assays at 9.1. karl fischer titration, and most certificates omit it, ymmv 19:21
  • chain_of_custody — the related substances table is the interesting part. a row of "not detected" means they did not look hard 19:36
BW

if the certificate date precedes the lot date, ask. sometimes it is sloppiness and sometimes it is not

BW

net peptide is gross mass minus water minus counterion. that is why 11.4mg of powder can be a correct 10mg vial

TW

single wavelength detection can miss a co-eluting impurity that does not absorb there

LL

not evidence on its own

medutest-report-a-2601.pdf
3 pages · 555 KB · not retained in the public archive
LL

no lot no certificate

LineClaimsDoes not claim
Purity (area %)fraction of detected material that is the targethow many mg are in the vial
Assay / contentmg of target per vialthat it is sterile
Water contentmass fraction that is wateranything about purity
Counterionsalt form and its mass sharenet peptide unless stated
Related substanceswhat else was seenwhat was not looked for
TW

the three way match is vial, certificate, invoice. if all three agree you have something

LL

if you only read one line on a certificate, read the lot number. if you read two, read the assay, take that with a pinch of salt

📉1🤝6
EO

an identical chromatogram on two different lots is the single biggest tell there is

*Medutest not the other one

CO

tfa counterion runs higher by mass than acetate. if the certificate does not say, you cannot do the mass balance

EO

update from 25 months ago: started asking every supplier for water content and about half now include it

BW

no lot number, no certificate. a document you cannot tie to the vial in your hand is decoration

certificate is dated before the lot date, is that just sloppiness

CO

water content is why an honest 10mg vial assays at 9.1. karl fischer titration, and most certificates omit it, ymmv

❤️4

is sterility tested as part of a normal purity certificate

a certificate should name the method, the instrument, the column and the operator. most name none of those

the related substances table is the interesting part. a row of "not detected" means they did not look hard

❤️1📉7👍1
LL

the retention time on this one is different from the last lot, problem

CO

retention time shifts between runs are normal. relative retention time is the number to compare

LL

should a certificate say which instrument and method they used