#hplc-massspec 2024-09-29

Sunday17 messages5 participantstimes are UTC
Highlights from this day
  • quiet.hours — sorry to jump in is monoisotopic or average mass the one on most certificates 21:41
  • area_percent — a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you 21:44
  • quiet.hours — do you trust an area percent from a shallow gradient ms/ms sequencing is the only thing i would call identity. everything else is circumstantial 21:49
  • deamidation — two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly if the number you are staring at is a third of what… 21:52
  • area_percent — unrelated but i was wrong about tailing factor. it is usually the injection or the pH, not the column 22:23
AP

tirzepatide is about 4813.5 Da, so 2+ is near 2407.8 and 3+ near 1605.5

QH

sorry to jump in is monoisotopic or average mass the one on most certificates

🙏16🧊10🤝18
AP

a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you

do you trust an area percent from a shallow gradient
ms/ms sequencing is the only thing i would call identity. everything else is circumstantial

DE

two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly
if the number you are staring at is a third of what you expected, do the charge state arithmetic before you panic

💀1👍13

unrelated but i was wrong about tailing factor. it is usually the injection or the pH, not the column

⚠️2📈6

is a tailing factor of anything specific a problem

DE

is there any point running a blank if the sample is the question

HH

relative retention time against a standard is comparable between labs. absolute minutes are not