Highlights from this day
  • half_life_hal — semaglutide is about 4113.6 Da. [M+2H]2+ lands near 2057.8, [M+3H]3+ near 1372.2 14:16
  • half_life_hal — truncations show as a mass deficit matching one or more residues. you need the sequence to say which 14:37
  • non_scale_win — has anyone ever had a result where the mass was right and the purity was wrong 16:24
HL

semaglutide is about 4113.6 Da. [M+2H]2+ lands near 2057.8, [M+3H]3+ near 1372.2

📉6⚠️12

truncations show as a mass deficit matching one or more residues. you need the sequence to say which

🧊14

integration difference

ask for the trace

a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you

shallow gradients hide related substances behind the main peak. a steeper one separates and looks worse, honestly

within inter-lab range

MM

a shoulder on a peak is information, and usually bad information, ymmv

MM

how much sample does a lab actually need for identity as well as purity

NS

a certificate with a number and no chromatogram is asking you to trust the integration you cannot see

run a blank

[edited]

has anyone ever had a result where the mass was right and the purity was wrong

fridge-thermometer.png
800 × 800 · 728 KB · not retained in the public archive

deamidation is about plus 0.98 Da. it is not a rounding error, it is a degradant

thats co-elution