- monoisotopic — if the number you are staring at is a third of what you expected, do the charge state arithmetic before you panic 14:24
- monoisotopic — the honest position is that hobby-level interpretation of a chromatogram is worth something but not much 14:42
- fridge_thermo — i asked VendorInvestigate for the raw trace and they sent it. that is the right answer to that question 15:11
- snac_and_water — update on the earlier thing deamidation is about plus 0.98 Da. it is not a rounding error, it is a degradant 15:26
- fridge_thermo — a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you, for what its worth 16:31
if the number you are staring at is a third of what you expected, do the charge state arithmetic before you panic
monoisotopic and average diverge more the bigger the molecule. say which one you are quoting
the honest position is that hobby-level interpretation of a chromatogram is worth something but not much
show me the spectrum
i asked VendorInvestigate for the raw trace and they sent it. that is the right answer to that question
update on the earlier thing deamidation is about plus 0.98 Da. it is not a rounding error, it is a degradant
plus 0.98 is deamidation
need more sample
semaglutide is about 4113.6 Da. [M+2H]2+ lands near 2057.8, [M+3H]3+ near 1372.2
a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you, for what its worth
why do two labs integrate the same peak differently
the first peak is the solvent. it is always the solvent