214nm sees the amide backbone so it sees everything. 280 only sees aromatics
[edited]#hplc-massspec 2026-01-24
- two_four_ceiling — is there any point running a blank if the sample is the question truncations show as a mass deficit matching one or more residues. you need the sequence to say which 18:56
- tesa_tess — why does integration change the number so much 19:37
- tesa_tess — update from 1 months ago: switched to asking for the chromatogram rather than the number and it changed everything, your mileage will differ 20:35
- seven_five_sweet — does retention time on its own tell you what the compound is update from 5 months ago: switched to asking for the chromatogram rather than the number and it changed… 20:40
- still_here_2024 — oxidation is about plus 15.99 Da and is usually methionine or tryptophan 20:53
thats a charge state
is there any point running a blank if the sample is the question
truncations show as a mass deficit matching one or more residues. you need the sequence to say which
can you tell a dimer from the chromatogram alone
if somebody says they ran a mass spec and does not show you a spectrum, that is a claim, not a result
relative retention time against a standard is comparable between labs. absolute minutes are not
deamidation is about plus 0.98 Da. it is not a rounding error, it is a degradant
tirzepatide is about 4813.5 Da, so 2+ is near 2407.8 and 3+ near 1605.5, thats just me
why does integration change the number so much
ms/ms sequencing is the only thing i would call identity. everything else is circumstantial, n of 1 obviously
how do you spot a deletion in a sequence from mass alone
ask for the trace
thats the solvent peak
update from 1 months ago: switched to asking for the chromatogram rather than the number and it changed everything, your mileage will differ
plus sixteen, is that always oxidation
has anyone ever had a result where the mass was right and the purity was wrong
does retention time on its own tell you what the compound is
update from 5 months ago: switched to asking for the chromatogram rather than the number and it changed everything
*Medutest not the other one
oxidation is about plus 15.99 Da and is usually methionine or tryptophan
monoisotopic and average diverge more the bigger the molecule. say which one you are quoting, take that with a pinch of salt
do the arithmetic
need more sample
why do two labs integrate the same peak differently