right so two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly
#hplc-massspec 2026-04-05
- VialBot — Archive lookup: quiet.hours first appears in this channel on 2025-08-23. 15:40
- otto_swirls — the honest position is that hobby-level interpretation of a chromatogram is worth something but not much 17:31
- peak_split — i have a number at about a third of what i expected, is that a charge state 17:33
- otto_swirls — how do you spot a deletion in a sequence from mass alone 18:16
tirzepatide is about 4813.5 Da, so 2+ is near 2407.8 and 3+ near 1605.5
genuine question whats a normal gradient for a peptide this size
a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you
whats the void volume peak, is that always solvent
shallow gradient
ms/ms sequencing is the only thing i would call identity. everything else is circumstantial
i was wrong about tailing factor. it is usually the injection or the pH, not the column
if somebody says they ran a mass spec and does not show you a spectrum, that is a claim, not a result
truncations show as a mass deficit matching one or more residues. you need the sequence to say which
has anyone ever had a result where the mass was right and the purity was wrong
the first peak is the solvent. it is always the solvent, ill dig out the number
do you trust an area percent from a shallow gradient
is monoisotopic or average mass the one on most certificates
co-elution is the failure mode that a beautiful single peak is most likely to be hiding
sorry to jump in you need more sample for identity than for purity. plan for that when you post it
is there any point running a blank if the sample is the question
Archive lookup: quiet.hours first appears in this channel on 2025-08-23.
a warm transit can put a visible degradant peak on a chromatogram. that is what 23 days at ambient does
need more sample
update from 3 months ago: switched to asking for the chromatogram rather than the number and it changed everything
plus sixteen, is that always oxidation
slightly off topic but coming back after 18 months, what changed in how you read these
how do you tell co-elution from a clean single peak
how much sample does a lab actually need for identity as well as purity
the honest position is that hobby-level interpretation of a chromatogram is worth something but not much
i have a number at about a third of what i expected, is that a charge state
how do you spot a deletion in a sequence from mass alone