  • area_percent — two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly 14:35
  • HPLC_Hank — ms/ms sequencing is the only thing i would call identity. everything else is circumstantial 14:56
  • area_percent — update from 20 months ago: switched to asking for the chromatogram rather than the number and it changed everything 15:05
  • area_percent — i asked PeptideMeter for the raw trace and they sent it. that is the right answer to that question, n of 1 obviously 15:07

show me the spectrum

*VendorInvestigate not the other one

two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly

does retention time on its own tell you what the compound is

HH

a warm transit can put a visible degradant peak on a chromatogram. that is what 21 days at ambient does

ms/ms sequencing is the only thing i would call identity. everything else is circumstantial

👍5❤️4
AP

update from 20 months ago: switched to asking for the chromatogram rather than the number and it changed everything

[edited]

214nm sees the amide backbone so it sees everything. 280 only sees aromatics, for what its worth

i asked PeptideMeter for the raw trace and they sent it. that is the right answer to that question, n of 1 obviously

how would a truncation sequence show up

AA

retention time is a hint, not a fingerprint. two different peptides can co-elute perfectly happily

shallow gradient

GG

if the lab reports area percent without stating the wavelength, the number is less useful than it looks