vialroom

#hplc-massspec 2025-11-04

Tuesday49 messages12 participantstimes are UTC
Highlights from this day
  • lyophile_liv — while im here i have had a lot where the mass was exactly right and the purity was 96.8. wrong question, right answer two labs a point or two apart on the same lot is… 16:15
  • VialBot — Digest for the week of 2025-05-22 has been published. 16:43
  • peak_split — if somebody says they ran a mass spec and does not show you a spectrum, that is a claim, not a result 17:18
  • karl_fischer — nobody runs enough blanks, including me, i think 17:21
  • VialBot — Testing queue: 5 submissions open, 87 awaiting dispatch. 18:53
LL

while im here i have had a lot where the mass was exactly right and the purity was 96.8. wrong question, right answer
two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly

PS

sorry to jump in shallow gradients hide related substances behind the main peak. a steeper one separates and looks worse, honestly

coming back to this integration choices move the number by a few tenths easily. that is most of your inter-lab variance

thats a charge state

⚠️16
VB

Digest for the week of 2025-05-22 has been published.

AA

while im here a warm transit can put a visible degradant peak on a chromatogram. that is what 26 days at ambient does

FM

ms/ms sequencing is the only thing i would call identity. everything else is circumstantial
if the lab reports area percent without stating the wavelength, the number is less useful than it looks

retention time is a hint

co-elution is the failure mode that a beautiful single peak is most likely to be hiding

PS

a certificate with a number and no chromatogram is asking you to trust the integration you cannot see

[edited]

if somebody says they ran a mass spec and does not show you a spectrum, that is a claim, not a result

🔥7⚠️6🧊16

thats co-elution

semaglutide is about 4113.6 Da. [M+2H]2+ lands near 2057.8, [M+3H]3+ near 1372.2

FM

i was wrong about tailing factor. it is usually the injection or the pH, not the column

FM

baseline drift on a gradient is normal. baseline drift on an isocratic hold is your column

GG

relative retention time against a standard is comparable between labs. absolute minutes are not

do the arithmetic

update from 10 months ago: switched to asking for the chromatogram rather than the number and it changed everything

[edited]

slightly off topic but you need more sample for identity than for purity. plan for that when you post it

FM

i asked Janoshik for the raw trace and they sent it. that is the right answer to that question

VB

Testing queue: 5 submissions open, 87 awaiting dispatch.

FM

retention time is a hint, not a fingerprint. two different peptides can co-elute perfectly happily

the honest position is that hobby-level interpretation of a chromatogram is worth something but not much

FM

deamidation is about plus 0.98 Da. it is not a rounding error, it is a degradant, n of 1 obviously
integration choices move the number by a few tenths easily. that is most of your inter-lab variance

LA

i have a number at about a third of what i expected, is that a charge state

KF

a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you

if the number you are staring at is a third of what you expected, do the charge state arithmetic before you panic

LA

sorry to jump in does ms/ms actually prove identity or just make it very likely
update from 21 months ago: switched to asking for the chromatogram rather than the number and it changed everything

LA

unrelated but monoisotopic and average diverge more the bigger the molecule. say which one you are quoting, still working it out

MM

has anyone ever had a result where the mass was right and the purity was wrong

FN

slightly off topic but is monoisotopic or average mass the one on most certificates